Re: absorbtion by air
Stanislaw Gierlotka ( xray@iris.unipress.waw.pl )
Thu, 19 Dec 1996 21:09:53 +0100 (MET)
I did that once and got some 30% more intensity with SIEMENS D5000.
You don't need to put the diffractometer into the closed bag - remember
that He is lighter than air. I just biult a sort of a tent on a wire frame
and supplied a He flow from the bottom. The problem might be with cooling
of the stepper motors. There will be no gas flow under the tent.
Actually that was why I could no continue working this way with my
SIEMENS. It uses a fan which cools the inside of the diffractometer.
I disconnected it for the time of testing but didn't dare to keep it
this way for long.
Stan
--------------------------+---------------------------------------
Stan(islaw) Gierlotka     |  High Pressure Research Centre
xray@iris.unipress.waw.pl |  Polish Academy of Sciences
X-Ray Diffraction Lab.    |  Sokolowska 29, 01-142 Warszawa, Poland
phone +(48 22) 6328497    |  phone (international) +(48) 3912 3276
 |                        |  phone (local)         +(48 22) 6325010
 |      |                 |  fax (international)   +(48) 3912 0331
 |      |   |       |     |  fax (local)           +(48 22) 6324218
_!_!____!___!_!__!__!!____!________________________________________